HOME SCREEN

The home screen displays
operating parameters
and information for the
entire Impella
System.
®
Display Element
1
Alarm window
2
Catheter serial number
3
System date and time
4
AIC serial number and
software version
5
System power area
6
Mute alarm indicator
6
Description
The alarm window displays up to 3 alarms simultaneously, in order of priority from top to bottom.
For each alarm, the alarm window displays:
• Alarm header – displayed in the left column; window is color-coded red for critical alarms, yellow for serious alarms, white
for advisory notifications, gray for resolved alarms
• Detailed text – up to 3 lines of instructions for resolving the alarm condition are displayed in the right column of the alarm
window next to the alarm header and subhead information
Displayed in the upper left of the display screen if a catheter is connected to the controller.
The current date (YYYY-MM-DD) and time (24-hour format; HH:MM) are displayed in the upper center of the screen display. (In
this example it is 2019 April 5 at 3:56 pm.)
The AIC serial number and current software version running on the console are displayed in the upper right hand corner of the
display screen.
System power information is located at the top of the display screen, to the right.
Battery status – bar within battery symbol indicates the overall remaining capacity of the batteries
• Full green bar for fully charged battery
• Partial green bar for battery that is at least 50% charged
• Partial yellow bar for battery that is between 16% and 50% charged
• Partial red bar for battery that is less than or equal to 15% charged
• Moving gray bar for battery that is in charging mode
• Numeric percentage of battery power remaining displayed to the left of the battery icon AC plug indicator
AC plug indicator
• White plug indicates that the controller is running on AC power
• White plug with a red X indicates no AC power detected and the controller is running on battery power
Displayed in place of the words "MUTE ALARM" when an alarm is silenced.
• Yellow bell with red X displayed when an alarm is muted
• Not displayed when an alarm is active (but not muted) or when there are no active alarms