Figure 5. Impella Catheter Inlet in Aorta (TEE)
• Catheter inlet area in aorta or near the aortic valve
• Catheter pigtail too close to the mitral valve
Figure 6. Impella Catheter in Papillary Muscle (TEE)
• Catheter pigtail in papillary muscle
• Catheter inlet area more than 4 cm below the aortic valve or lodged
between papillary muscle and the myocardial wall
• Catheter outlet area too close to the aortic valve
Four Transthoracic Echocardiographic (TTE) images
Figure 7. Correct Impella Catheter Position (TTE)
• Catheter inlet area 3.5 cm below the aortic valve
• Catheter outlet area well above the aortic valve
• Catheter angled toward the left ventricular apex away from the heart
wall and not curled up or blocking the mitral valve
Automated Impella Controller
Figure 8. Impella Catheter Too Far into Left Ventricle (TTE)
• Catheter inlet area more than 4 cm below the aortic valve
• Catheter outlet area across or near the aortic valve
Figure 9. Impella Catheter Inlet in Aorta (TTE)
• Catheter inlet area in aorta or near the aortic valve
• Catheter pigtail too close to the mitral valve
Papillary muscle
Figure 10. Impella Catheter in Papillary Muscle (TTE)
• Catheter pigtail in papillary muscle
• Catheter inlet area more than 4 cm below the aortic valve or lodged
between papillary muscle and the myocardial wall
• Catheter outlet area too close to the aortic valve
Aorta
23