Adding A Spirometry Maneuver; Conducting A 'Post' Test; Saving And Retrieving Measurements Directly In Easyone; Quick Test - NDD EasyOne-CS Manuel De L'utilisateur

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 106
2030434-002 Rev. B

5.3.2 Adding a spirometry maneuver

If you wish to later add spirometry maneuvers to a test session, for example if the patient was ex-
hausted or if you wanted to test other patients in the meantime, please proceed as follows:
Select "Perform Test" in the main menu
Choose the field RECALL and press ENTER
Scroll through the list of tests until you get to the desired test and press ENTER
Select the field ADD
Perform the additional spirometry maneuvers
Please note that it is only possible to add a trial to a previous test that was performed on the same day.
No spirometry maneuvers can be added to tests that were generated with CardioSoft

5.3.3 Conducting a 'Post' test

In order to add a Post test after an FVC or FVL test, click on the POST button on the Result screen.
Proceed as follows if you wish to conduct a Post test from the main menu:
Select "Perform Test" in the main menu
Choose the field RECALL and press ENTER
Scroll through the list of tests until you get to the desired pre-med test and press ENTER
Select the field POST
Now proceed as described in Chapter 5.3.1.

5.3.4 Saving and retrieving measurements directly in EasyOne

EasyOne-CS saves all test results automatically to the spirometer's internal memory. No data can be
lost even if the batteries are removed from the instrument. If the internal memory of the EasyOne-CS is
full (after approx. 700 normal spirometry measurements), the oldest test in the internal memory of the
instrument is overwritten
You can recall saved measurements for the purpose of conducting a new test with the same patient,
adding maneuvers, adding a post-test, viewing results, or printing results.
You can add a maneuver or a post-test only on the same day that the original test was performed.
Please read Chapter 6.5 und 5.3.2.
In order to add a test to an old measurement, choose "Perform Test" in the main menu and then
choose RECALL. Follow the rest of the instructions.
In order to view an old test, choose "View Results" in the main menu and choose the desired test.

5.3.5 Quick Test

You have the option to perform a quick test without entering patient data. Select "Perform Test" in the
main menu and then QUICK. Choose the desired test using the arrow keys (<) or (>) and perform the test.
Please note:
When Quick Test is selected, no comparison to predicted normals are displayed
or printed. Predicted normals are only available when age, height and gender
are entered.
It is possible to enter patient data after having performed a quick test. Proceed as described in Chapter
5.3.6. Once patient data is entered predicted normals will be displayed and printed.
2
No data is deleted or overwritten when synchronizing with the PC. Overwriting tests relates only to test data in the EasyOne-CS's internal