Technical Data

Power connection:
Frequency:
Max. power consumption:
Max. current consumption:
Protection class:
Protection level:
Moisture protection:
Dimensions:
Weight:
Operating conditions:
Storage and transportation conditions:
Manufactured and tested according to:
EMC:
Insufflation medium:
Maximum output pressure:
Maximum gas supply pressure:
Minimum gas supply pressure (bottle):
Minimum gas supply pressure (house gas):
Maximum gas flow:
Pressure range:
Pressure display resolution:
Gas flow display resolution:
Volume display resolution:
Connections/ports:
100-240 V~
Mains fuse T 3.15 A H 250 V
Connection for potential equalization
50-60 Hz
150 VA
100 V~ 60 Hz: 1.5 A (US, UK, CA: 1.5 A)
240 V~ 50 Hz: 0.63 A (US, UK, CA: 0.63 A)
I
Type BF
IP21
Width x Height x Depth
406 x 150 x 395 [mm]
approx. 10 kg/22 lbs
10-40 °C/50-104 °F
30-75% rel. humidity
700 - 1060 hPa ambient pressure
maximum usage altitude: 3000 m above sea level
-20 - +60° C /-4 - +140° F
10-85% rel. humidity
IEC 60601-1:2005+Corr2006+Corr2007+Int1:2008+Int2:2009
EN 60601-1-2:2007
Medical grade CO
2
75 mmHg (1 mmHg = 1.33 mbar = 133 Pa)
80 bar/1160 PSI
15 bar/218 PSI
3.4 bar/50 PSI
High Flow: 40 l/min
Bariatric: see equipment data label on the back of the device
1-30 mmHg
1 mmHg
1 l/min
in the range of 0 to 9.9 l resolution = 0.1 l
in the range of 10 to 999 l resolution = 1 l
Service interface, interface Arthrex control bus (ACB)
Technical Data
EN
61