Insufflation: High Flow
While Insufflation: High Flow any pressure loss due to leaks can be quickly equal-
ized. APC technology (Advanced Pressure Control) enables raising the actual
pressure gently to the level of the nominal pressure. In case of large volumes, the
actual pressure usually does not exceed the nominal pressure (see chapter 9
Safety Functions, page 42).
Start the device by pressing the START function field.
Veress Insufflation is displayed with a nominal flow setting of < 6 l/min. Above
5 l/min Insufflation: High Flow) is displayed.
Stop the device by pressing the STOP function field.
NOTE!
Tube, filter, and instrument can reduce the gas flow.
6.5

Performing the Function Test in High Flow Operating Mode

Sterilize reusable instruments and tubing before surgery to prevent infections.
Check all the single-use/disposable items before removing them from the pack-
age to ensure that the packaging is intact and that the expiration date is still val-
id.
For your own safety and that of your patient, use only original accessories.
WARNING!
The function test must be performed prior to each surgical procedure.
(1) (2) (3)
(4)
(8)
(7)
1. The device is switched off, no tube set is connected.
2. Make sure the gas supply is connected and open.
3. Use the ON/OFF switch (1) to turn the device on. The device now conducts a
(5)
(6)
Device Use and Control with High Flow Operating Mode
Preparation
Testing the device
Fig. 6-1
Layout for testing the device
ON/OFF switch
(1)
Nominal pressure display
(2)
(3)
START/STOP function field
Gas consumption display
(4)
Nominal flow display
(5)
Insufflation tube set
(6)
Stopcock (valve)
(7)
Veress cannula
(8)
EN
23