CAUTION!
Please note that with flow values up to 5 l/min the Veress insufflation operating
mode is automatically activated (the venting valve is switched off). The gas
transport is also specifically customized for use with a Veress cannula.
WARNING!
This manual does not include instructions for the safe use of the Veress cannula.
Only after you have verified that an aeroperitoneum can be generated should a
gas flow of more than 3 l/min and a pressure of more than 10 mmHg be selected.
Insert the Veress cannula into the abdomen. Check to see if the Veress cannula is
correctly positioned in the abdomen.
1. Attach the insufflation tube to the Veress cannula.
2. Select the desired nominal pressure and nominal gas flow.
3. Start insufflation: Press the [START] function field.
Check the actual pressure display and the gas consumption display.
7.6.2

Insufflating with the Trocar

1. Insert the trocar into the abdomen.
2. Connect the Luer Lock connection of the insufflation tube to the trocar.
3. Make sure the trocar is correctly positioned in the abdomen. Then select the
desired pressure and desired gas flow as intraoperative conditions.
4. The actual pressure display shows the current measured value for insuffla-
tion. As soon as this value approximates the selected nominal value, the gas
flow is automatically minimized. The gas consumption display shows the vol-
ume of gas consumed.
5. Check how the patient's body reacts to the selected pressure and gas flow
rate. Compare the abdominal filling rate to the selected nominal pressure.
You can change the nominal gas flow and the nominal pressure during sur-
gery without interrupting the insufflation process. The Luer Lock connection
of the measuring tube can be connected to another trocar that also inserted
when using a tube set with measuring line. This enables the continuous mea-
suring of the pressure.
7.7

Stopping Insufflation

1. Press the [STOP] function field. The following values are displayed:
• Gas consumption display: last measured value.
• Actual pressure: Current measured value
• Actual gas flow: 0 l/min
• Nominal pressure: last set value In case of exceeding the safety limit, the
nominal pressure value will be reset to the safety limit value.
• Nominal gas flow: last set value.
The display status field depicts Insufflation Stopped with Push
2. Remove the tube set from the device. Observe applicable hygiene regulations
when disposing of the tube set.
CAUTION!
If the tube set remains connected to the device, there is the danger that leftover
fluid in the tube or the instruments will be sucked into the device.
3. Close the gas supply.
4. Use the ON/OFF switch to turn the device off.
NOTE!
Observe applicable hygiene regulations when disposing of the tube set.