Error and Warning Messages

Error and warning messages
Check gas supply
Change gas bottle!
Overpressure
Venting system active
Overpressure/venting system
active.
Occlusion
Contamination
Contamination/Call service
Gas heater defective/Call service
Cause
(During device check)
Existing gas supply pressure is
too low.
(During surgery)
• The gas supply pressure has
dropped below 15 bar.
• Insufficient house gas supply. • Check the house gas supply.
The gas supply pressure has
dropped below 30 bar.
The pressure monitor shows
that the actual pressure is at
least 4 mm Hg above the nomi-
nal pressure.
The actual pressure has reached
30 mmHg/20 mmHg (depend-
ing on the operating mode).
The actual pressure is at least
2-5 mmHg and 2-5 s above the
nominal pressure.
The pressure monitor indicates
that the actual pressure is
exceeding the nominal pressure
by 2-5 mmHg and for longer
than 2-5 seconds. The overpres-
sure was not reduced within 5
seconds by the venting system.
Tube or instrument occlusion.
Faulty Veress needle insertion.
Stopcock is closed.
Fluid has penetrated the device
through the patient gas outlet.
The device is contaminated
with fluid.
Gas heater malfunction.
Troubleshooting
• Open gas bottle or
• Replace the gas bottle.
• Check the house gas supply
• Detach the insufflation tube.
• Close gas supply valve.
• Replace gas supply.
• Open gas supply valve.
• Connect insufflation tube.
• Continue insufflation.
• Open or remove pressure reduction valve if in line
with gas bottle.
• Check if appropriate house gas connector is used.
• Prepare for changing the gas bottle.
• Determine the cause for exceeding the nominal
pressure. Check the electronic controls of the device
if overpressure exists for a longer period of time
(see chapter 11 Annual Inspection, page 50).
• Reduce the nominal pressure and determine the
cause of exceeding the nominal pressure.
• Determine the cause for exceeding the nominal
pressure. Check the electronic controls of the device
if overpressure exists for a longer period of time
(see chapter 11 Annual Inspection, page 50).
• Determine the cause for exceeding the nominal
pressure. Check the electronic controls of the device
if overpressure exists for a longer period of time
(see chapter 11 Annual Inspection, page 50).
Reduce the nominal pressure. Check if the instru-
ment's stopcock is open or the tube is plugged up.
• Localize the cause and open/eliminate the occlu-
sion.
• Check that the Veress cannula is positioned cor-
rectly in the abdomen and make sure the instru-
ment's stopcock is open.
• The message is repeated with each Start/Stop. It is
possible to continue using the device with this error
message until the device is turned off with the ON/
OFF key.
• The device must be checked by an authorized ser-
vice technician or clearly marked with a label refer-
ring to the contamination and then twice enclosed
in a safety foil, sealed, and returned to the manu-
facturer for repairs.
• Check gas heating with a new tube. If this error
message is displayed again, have the gas heater
checked by a qualified service technician. It is possi-
ble to continue using the device without gas heater.
Error and Warning Messages
EN
59