Selecting High Flow Insufflation Mode - Arthrex Synergy Insufflation FM134 Instructions D'utilisation

Insufflateur pour la laparoscopie
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 141
WARNING!
Dehydration
Insufflation can lead to dehydration of the tissue. This can result in organ tissue
damage and cardiovascular reactions of the patient. Long surgeries and large
leaks increase the risk of dehydration (especially at the insertion points of the
trocars or when changing instruments).
WARNING!
Embolism/insufflation of internal organs
Improper placement of the insufflation instrument could cause gas penetrating
a vessel or an internal organ, resulting in air or CO
risk, use a low flow rate for the first insufflation and ensure that the insufflation
instrument is correctly positioned. Check the position of the insufflation instru-
ment immediately if the actual pressure rapidly reaches the nominal pressure
value. CO
embolisms can also be caused by a high intra-abdominal pressure.
2
Avoid high-pressure settings and close damaged blood vessels at once.
WARNING!
Additional insufflation sources
The use of additional insufflation sources increases the intra-abdominal pres-
sure. Continuously monitor intra-abdominal pressure over the course of the en-
tire insufflation if additional sources are used.
WARNING!
Please read the device-specific risks and warnings in chapter 3.3 General Warn-
ings and Precautions, page 5.
6.2

Selecting High Flow Insufflation Mode

1. Selection of the insufflation mode can be done only while insufflation is
stopped. If the High Flow operating mode is not yet set, press the function
field Select Insufflation Mode for 2 seconds to load the insufflation overview.
2. The display depicts the different insufflation operating modes that can be ac-
tivated. Select the desired insufflation operating mode by pressing the High
Flow function field.
3. The procedure profile is depicted on the display. The displayed parameters
correspond with the factory settings or the values set in the configuration
menu (see chapter 8.1 Configuration Menu I, page 36).
Device Use and Control with High Flow Operating Mode
embolisms. To reduce the
2
(1)
EN
21