Care and Maintenance
EN
Fig. 10-1 Opening the fuse holder
46
10.4

Replacing the Fuse

CAUTION!
Before replacing the fuse, check the values for the fuse to be inserted in accor-
dance with chapter 14 Technical Data, page 61.
The fuse may be defective and is in need of replacement if:
• Indicators and display (if present on the device) do not light up,
• the device does not function.
Check whether
• the main power supply cable is properly connected to the power supply input
and to a grounded safety wall socket,
• the house power supply fuse is functioning.
WARNING!
Unplug the power cable from the device before checking the fuse.
The device does not have to be opened to replace the fuse.
1. Switch device off.
2. Disconnect device from power supply by pulling the mains plug from the
mains socket.
3. Remove the power connection cable from mains socket.
4. The fuse holder is located next to the mains socket.
5. Remove the fuse holder as described in Fig. 10-1 Opening the fuse holder .
6. A Undo the latch of the fuse holder with two fingers.
7. B Remove the fuse holder.
8. C Check fuse.
9. Insert a new fuse. Use only the specified type of fuse (see chapter 14 Techni-
cal Data, page 61).
10. Insert the fuse holder until it can be heard snapping into place.
11. Use the power cable to reconnect the shockproof safety socket with the rear
mains socket.
A
B
C