WARNING!
If the temperature sensor measures a gas temperature > 42° C, unplug the heat-
ing plug from the device. Hot gas in the abdomen can lead to serious injuries.
A warning signal is audible. Gas heating and insufflation are deactivated.
1. Disconnect the insufflation tube from the trocar or Veress needle.
2. Unplug the plug for the gas heating from the device.
3. Press the Start/Stop key. The device insufflates without heating the gas.
4. Let hot gas escape until the tube is only warm to the touch.
5. Continue operation without connecting gas heating.
6. Check gas heating after surgery using a different tube. Turn device off and on
again. Gas heating is reactivated.
7. Should the error message be displayed again, you can continue using the de-
vice without gas heating by observing the risks for hypothermia.
8. Call an authorized service technician to check the gas heating.
Operating the Device - General