Device Use and Control with High Flow Operating Mode
EN
Testing gas heating functionality
24
device check.
4. Select High Insufflation operating mode in the insufflation mode selection.
5. Connect an original insufflation tube set (6) to the device.
6. Attach the insufflation tube to the Veress cannula (8).
7. Attach the stopcock (valve) (7) to the Veress cannula.
8. If the gas consumption display function field does depict a value, press the
function field to reset the display to (4) zero 0.
9. Set the nominal pressure to 15 mmHg (2) and the gas flow to 3 l/min(5).
10. Start insufflation: Press the [START] (3) function field.
11. The display status line depicts Occlusion after max. 4 seconds.
12. Stop insufflation: Press the [STOP] (3) function field.
WARNING!
If the actual gas consumption is higher than 0.4 l, there is a leak in the system. If
this happens, use steps 13 to 15 outlined below to locate the leak.
13. Repeat items 7 to 11 without Veress cannula and with closed tube end. The
previously connected Veress cannula has a leak if gas consumption is now be-
low 0.4 l.
14. Repeat items 8 to 12 without Veress cannula and without tube if another leak
becomes apparent. Close the end of the insufflation tube connection for this
test. If the gas consumption is then below 0.4 l, the previously used tube set
has a leak.
15. If another leak is detected, this leak is then directly within the device. Make
sure the device can no longer be operated until an authorized service techni-
cian conducts the appropriate checks.
WARNING!
Never work with a leaky insufflation tube, accessory, and/or device. This can
lead to an incorrect measurement of the actual pressure values, which can cause
an uncontrolled pressure increase in the abdomen.
The gas heating function will be indicated with the gas heating symbol on the
display if the "heating tube set" is used.
6.5.1

Filling Tube System with CO2

At least 1 l of CO
must be released from the system before every surgery while
2
the tube is connected and the end of the tube is open to expel any air within the
tube system and the device.
1. Start insufflation: Press the [START] function field.
Wait until the gas consumption display shows 1.0 l.
2. Stop insufflation: Press the [STOP] function field.
3. Press the function field to reset the gas consumption display (see Fig. 5-3
Screen displays, (13)), to reset the gas consumption display to 0.0 liters. This
ensures the gas consumption is correctly displayed during surgery.
The function test is complete. The device is tested and ready to be used for sur-
gery.
WARNING!
Do not use this device if a defect is suspected or detected during the function
check. This also applies to obvious defects, especially defects on the power plug
and power cable.