Initial Device Startup

EN
Delivery inspection
Returning the device
Setting up the device
Mains connection
Grounding contact
Only for U.S. operators
12
4
Initial Device Startup
Always check all parts and accessories of the device immediately after receiving
the shipment. The manufacturer considers only replacement claims that have
been immediately submitted or reported to a sales representative or an autho-
rized service company.
4.1

Scope of Delivery

• Device Synergy Insufflation Type FM134
• Instructions for Use
• Power supply cord US, 2.5 m
Use the original packaging material if the device has to be returned. The manu-
facturer is not responsible for transport damages due to insufficient or unsuit-
able packaging.
Please make sure that all required information has been supplied:
• Name of owner
• Address of owner
• Device type and model
• Serial number of the equipment (see identification plate)
• Detailed description of defect
4.2

Preparing the Device

Place the device on a level surface and install in a dry environment. The ambient
temperature and humidity must meet the requirements mentioned in chapter
14 Technical Data, page 61.
WARNING!
Not explosion-proof
The device is not explosion-proof. Do not use in an area where flammable anes-
thetic gases are present.
WARNING!
Position the device in such a way that it is easy to operate and switch off.
CAUTION!
Check to make sure the available mains voltage matches the data listed on the
type label attached to the back of the device. Incorrect voltage can cause errors
and malfunctions and may destroy the device.
Make sure the connection data and technical specifications of the power supply
comply with DIN VDE or national requirements. The mains connection cable may
be plugged only into a properly installed, grounded safety wall socket (shock-
proof socket) (see DIN VDE 0100-710). Read the device label located in rear of de-
vice (type plate) to determine the operating voltage of the device.
The power connection must be equipped with a grounding contact. Use the orig-
inal power cable (if included in scope of delivery) to establish a connection be-
tween the mains wall socket and the non-heating device plug located in the rear
of the device.
Only use a certified (UL-listed), removable mains connection cable, type SJT, min-
imal 18 AWG, 3 leads. The plug connectors must comply with NEMA 5-15 or IEC