12.4
Guidelines and Manufacturer's Statement - Electromagnetic In-
terference Immunity
The insufflator is intended for use in an electromagnetic environment as de-
scribed below. The user/operator of the insufflator should make sure the device
is operated within such an environment.
Electromagnetic
Test level
interference immu-
nity tests
Discharge of static
± 6 kV contact
electricity (ESD)
discharge
according to
± 8 kV air dis-
IEC 61000-4-2
charge
Electrical fast tran-
± 2 kV for
sients / bursts
power lines,
according to
± 1 kV for
IEC 61000-4-4
input and out-
put lines
Surges according to
± 1 kV normal
IEC 61000-4-5
mode voltage,
± 2 kV com-
mon mode
voltage
Blackouts, brown-
< 5% U
T
outs, and fluctua-
95% dip in the
tions of the power
U
) for ½ cycle
T
supply according to
40% U
IEC 61000-4-11
T
dip in the U
for 5 cycles.
70% U
T
dip in the U
for 25 cycles.
< 5% U
T
95% dip in the
U
) for 5 s
T
Supply frequency
3 A/m
magnetic field (50/
60 Hz) according to
IEC 61000-4-8
*Note: U
is the mains alternating voltage before applying the test levels.
T
Compliance
Electromagnetic envi-
ronment guidelines
In compliance
Floors should be made
from wood or concrete
or covered with ceramic
tiles. If the floor cover-
ing consists of synthetic
material, the relative
humidity should be at
least 30%.
In compliance
The quality of the sup-
ply voltage should be
the same as the voltage
of a typical business or
hospital environment.
In compliance
The quality of the sup-
ply voltage should be
the same as the voltage
of a typical business or
hospital environment.
* (>
In compliance
The quality of the sup-
ply voltage should be
the same as the voltage
of a typical business or
(60%
hospital environment.
)
If the user/operator of
T
device requires the con-
tinuation of functional-
(30%
ity after power
)
T
interruptions/disrup-
tions, it is recom-
(>
mended to provide the
device with power from
an uninterruptible
power supply.
In compliance
Magnetic fields of the
mains power fre-
quency should comply
with the typical values
of business and hospi-
tal environments.
Electromagnetic Compatibility
55
EN