7.5
Performing the Function Test in Bariatric Operating Mode before
Using the Device during Surgery
Sterilize reusable instruments and tubing before surgery to prevent infections.
Check all the single-use/disposable items before removing them from the pack-
age to ensure that the packaging is intact and that the expiration date is still val-
id.
For your own safety and that of your patient, use only original accessories.
WARNING!
The function test must be performed prior to each surgical procedure.
(1) (2) (3)
(4)
(8)
(7)
1. The device is switched off, no tube set is connected.
2. Make sure the gas supply is connected and open.
3. Use the ON/OFF switch (1) to turn the device on. The device now conducts a
device check.
4. Select Bariatric operating mode in the insufflation mode selection.
5. Connect an original insufflation tube set (6) to the device.
6. Attach the insufflation tube to the Veress cannula (8).
7. Attach the stopcock (valve) (7) to the Veress cannula.
8. If the gas consumption display function field does depict a value, press the
function field to reset the display to (4) zero 0.
9. Set the nominal pressure to 15 mmHg (2) and the gas flow to 5 l/min (5).
10. Start insufflation: Press the [START] (3)function field.
11. Insufflate approx. 30 s. 30 seconds. The display status line depicts Occlusion
after max. 4 seconds.
12. Stop insufflation: Press the [STOP] (3) function field.
WARNING!
If the actual gas consumption is higher than 0.4 l, there is a leak in the system. If
this happens, use steps 13 to 15 outlined below to locate the leak.
13. Repeat items 7 to 11 without Veress cannula and with closed tube end. The
previously connected Veress cannula has a leak if gas consumption is now be-
low 0.4 l.
14. Repeat items 8 to 12 without Veress cannula and without tube if another leak
becomes apparent. Close the end of the insufflation tube connection for this
(5)
(6)
Device Use and Control with Bariatric Operating Mode
Preparation
Testing the device
Fig. 7-1
Layout for testing the device
ON/OFF switch
(1)
(2)
Nominal pressure display
START/STOP function field
(3)
Gas consumption display
(4)
Nominal flow display
(5)
Insufflation tube set
(6)
Stopcock (valve)
(7)
Veress cannula
(8)
EN