Switching The Device On; Insufflation Tube Connection - Arthrex Synergy Insufflation FM134 Instructions D'utilisation

Insufflateur pour la laparoscopie
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 141
House gas supply pressure OK
(green)
House gas supply pressure too low
(red)
5.4

Switching the Device On

1. Connect the device with the mains power.
2. Connect the gas supply to the gas connection port and open the gas supply.
3. Press the ON/OFF switch. The device turns on.
4. The device performs a device check after being switched on. The company
logo is displayed on the touch screen for 3 seconds after the successful device
test as well as the lines Synergy Insufflation -> Device Test -> Device OK. In
case the device check fails and an error message is displayed, please see chap-
ter 13 Error and Warning Messages for additional information.
5. The display depicts an insufflation operating modes overview. Press the re-
spective function field to select the corresponding desired operating mode
(e.g., High Flow).
5.4.1

Insufflation Tube Connection

1. Always connect a hydrophobic filter to the silicone tube or use the PVC tube
set with filter.
2. Connect the insufflation tube to the insufflation tube connection and to the
Veress cannula/trocar.
ISO 5356-1 connection
A male ISO connector is located at the device (ISO-M). Connect the female ISO
connector (ISO-F) of the filter or tube firmly with the ISO-M connector.
Operating the Device - General
(2)
(1)
(3)
EN
17

Publicité

Table des Matières
loading

Table des Matières