Setting The Gas Flow Rate; Setting The Maximum Nominal Pressure - Arthrex Synergy Insufflation FM134 Instructions D'utilisation

Insufflateur pour la laparoscopie
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 141
8.2
Configuration Menu II
While insufflation is stopped, press the Actual function field for 2 seconds to ac-
cess the Configuration menu. In the Configuration menu I, press the  to access
the Configuration menu II.
8.2.1

Setting the Gas Flow Rate

Rate
Operating
Default setting
mode
Flow rate 1 Bariatric
5 l/min
High Flow
3 l/min
Flow rate 2 Bariatric
25 l/min
High Flow
20 l/min
Flow rate 3 Bariatric
Max. value*
High Flow
40 l/min
**see equipment data label on the back of the device
In Configuration Menu II, press the Gas flow rates function field to access the se-
lection menu.
The display switches to the gas supply selection menu.
Press the function field (flow rate 1: 3 l/min), for example, to set gas flow rate 1.
The display switches to the setting menu for gas flow rate 1.
Press the º or » function field to set the value.
You can now do the following:
1. save the selected value by pressing the (SAVE) function field. The display de-
picts (SAVED) for 2 seconds. After saving you return automatically to the pre-
vious level.
2. press the  function field to return to the previous menu level without sav-
ing,
3. or press (EXIT) to return to the work screen without saving.
8.2.2

Setting the Maximum Nominal Pressure

Operating mode
Default setting
Bariatric
30 mmHg
High Flow
30 mmHg
In Configuration Menu I, press  to access Configuration Menu II.
The display changes to the Configuration menu II.
In Configuration Menu II, press the Maximum nominal pressure function field.
The display switches to the Maximum nominal pressure selection menu.
Press the º or » function field to set the value.
Configuration menu II
Range
1-20 l/min
1-20 l/min
Flow rate 1-max. value*
Flow rate 1-40 l/min
Flow rate 2-max. value*
Flow rate 2-40 l/min
Range
5-30 mmHg
5-30 mmHg
Configuration Menu (Overview)