Operating the Device - General
EN
Fig. 5-3
Screen displays
Insufflation mode selection
(1)
(2)
Insufflation mode display
Status display/error and warning
(3)
messages
Menu
(4)
Increasing nominal gas flow
(5)
Nominal gas flow display
(6)
Decrease nominal gas flow
(7)
Actual gas flow display
(8)
(9)
START/STOP function field
Gas heater connected/ready
(10)
Gas supply display
(11)
Gas consumption display
(12)
Function field to reset the gas con-
(13)
sumption display (reset button)
Decrease nominal pressure
(14)
Nominal pressure display
(15)
(16)
Increase nominal pressure
Abdominal pressure display
(17)
Gas supply displays
Gas supply with gas bottle
House gas supply
16
5.3

Touch Screen Display

(1)
(17)
(16)
(15)
(14)
(13)
(12)
(11)
The above depiction of the touch screen also shows all display and function
fields.
The status of the gas supply is monitored by the device and indicated with sym-
bols and acoustic signals (see chapter 9 Safety Functions for gas pressure display
information).
The following gas bottle pressures are displayed:
> 50 bar
40 - 50 bar
30 - 40 bar
15 - 30 bar; Three acoustics warning signals (beeps) can be heard
and the message "Change gas bottle" is displayed. User is advised to
prepare for changing the gas bottle.
< 15 bar; Three warning signals can be heard and the message
"Check gas supply" is displayed. Immediately replace the gas bottle.
If the gas cylinder pressure keeps falling, warnings will appear to remind the user
to replace the gas bottle immediately. At < 5 bar and again at 0 bar, 5 beeps will
sound and the display depicts "Check Gas Supply." Insufflation stops at 0 bar.
The following house gas supply pressures are displayed:
(2)
(3)
(4)
(10)
(9)
(5)
(6)
(7)
(8)