Configuration Menu (Overview)
EN
Functions of the configuration menu
Setting the venting valve status
36
8.1

Configuration Menu I

While insufflation is stopped, press the Menu function field (see Fig. 5-3 Screen
displays (4)) for 2 seconds to open Configuration Menu I.
Depiction/functions in configuration menu
Tap the "arrow forward" function field to access the next
menu on the same level.
Tap the "arrow backward" function field to access the
previous menu on the same level without saving set-
tings or to access the previous menu of the next higher
menu level.
Tap the (SAVE) function field to save settings. The display
depicts (SAVED) for 2 seconds. After saving you return
automatically to the previous level.
Save
Tap the (EXIT) function field to exit the menu or return to
the work screen.
Exit
8.1.1

Setting First Nominal Pressure

Operating mode
Default setting
Bariatric
15 mmHg
High Flow
15 mmHg
In Configuration Menu I, press the First nominal pressure function field to access
the setting.
The display switches to the First nominal pressure menu.
Press the function field º or » to set the First nominal pressure.
You can now do the following:
1. save the selected value by pressing the (SAVE) function field. The display de-
picts (SAVED) for 2 seconds. After saving you return automatically to the pre-
vious level.
2. press the  function field to return to the previous menu level without sav-
ing,
3. or press (EXIT) to return to the work screen without saving.
8.1.2

Setting the Venting Valve

In Configuration Menu I, press the Venting Valve function field to access the
venting system selection.
The display switches to the Venting Valve selection menu.
Factory settings
Operating
Default setting
mode
Bariatric
With veress insufflation OFF With veress insufflation OFF,
Range
1-15 mmHg
1-15 mmHg
Range
Venting system OFF