Electromagnetic Compatibility

EN
Precautionary measures
ESD (Electrostatic Discharge) precautionary
measures
Tube heating
Cable for centralized device control (ACB)
54
12
Electromagnetic Compatibility
Medical devices are subject to special safety and protective measures concerning
electromagnetic compatibility (hereafter abbreviated as EMC).
This device is to be used only for the purposes described in the manual and has
to be installed, set up, and operated in compliance with the EMC notes and in-
structions.
12.1

Impact of Mobile and Portable HF Communication Devices

The emission of high frequency energy by mobile communication devices may
impact the function of the electrical medical device. Operating such devices (e.g.,
cell phones, GSM phones) in the proximity of the electrical medical device is pro-
hibited.
12.2

Electrical Connections

Do not touch electrical connections identified with this warning label. Do not es-
tablish a connection between these plugs and sockets without first implement-
ing precautionary ESD (electrostatic discharge) measures.
The following are ESD precautionary measures:
• Apply potential equalization (PE), if available on your equipment, to all devices
to be connected.
• Use only the listed equipment and accessories.
Hospital employees should be informed about and trained in ESD precautionary
measures.
12.3

Accessories

The device can be used with a reusable or disposable tube set with gas heater val-
idated for this purpose.
Only a cable for the central unit control (ACB) may be connected to the device.