Devices to be Used within the Lumens of VivaSight 2 DLT
Ambu® VivaSight
TM
Suction Catheter
Other Devices
Other suction catheters
Bronchoscopes
Airway exchange catheters
Airway intubation catheters
Size guide for selection of appropriate size of devices to be used within the VivaSight 2 DLT lumens:
Tube Sizes
Ambu® VivaSight
2 DLT 35 Fr
TM
Ambu® VivaSight
2 DLT 37 Fr
TM
Ambu® VivaSight
2 DLT 39 Fr
TM
Ambu® VivaSight
2 DLT 41 Fr
TM
*The effective inside diameter is intended as a guide for selecting the appropriate diameter of a
bronchoscope or other device to be passed through the lumens. The effective inside diameter
differs from the ID by taking into account the restriction on the lumen at the camera position.
Ventilation Equipment
Lung ventilation systems with 15 mm female conical connectors in compliance with ISO 5356-1.
Accessories
Standard 6 % conical Luer syringes.
Manometer pressure gauges.
2.3. VivaSight 2 DLT Parts
14
15
16
17 18
19 20
8
9
10 11
No.
Part
1
Tracheal pilot balloon with
check valve
2
Flush port
3
Tracheal tube
REF Number:
412100000
Effective Inside
Diameter*
Bronchial [mm]
Min. 4.4
Min. 4.6
Min. 4.8
Min. 5.0
1
2
12
Function
Provides luer compatible port for tracheal cuff
inflation/deflation and indicates state of tracheal
cuff inflation/deflation
Provides luer compatible port for injection of air and
saline for cleaning of camera lens
Channels air for ventilation or deflation of right lung
Effective Inside Diameter*
Tracheal [mm]
Min. 4.4
Min. 4.6
Min. 4.8
Min. 5.0
3
4
5 6 7
13
7