Guidance and manufacturer's declaration – electromagnetic immunity
The system is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should assure that it is used in such an environment.
Immunity test
Conducted
Radio
Frequency
IEC 61000-4-6
Radiated Radio
Frequency
IEC 61000-4-3
14
IEC 60601-1-2
test level
For power lines
3 V RMS
0,15 MHz –
80 MHz
6 V RMS
in ISM bands
80 % AM at
1 kHz
3 V/m 80 MHz to
2.7 GHz
80 % AM at
1 kHz
Proximity to
RF wireless
communications
equipment
385MHz-
5785MHz,
up to 28V/m
Compliance
Level
For power lines
3 V RMS
0,15 MHz –
80 MHz
6 V RMS
in ISM bands
80 % AM at
1 kHz
3 V/m 80-2700
MHz
80 % AM at
1 kHz
Proximity to
RF wireless
communications
equipment
385MHz-
5785MHz,
up to 28V/m
Electromagnetic
Environment Guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the system,
including its cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
d = 1.17
d = 1.17
80 MHz to 800 MHz
d = 2.33
800 MHz to 2.7
GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey
a) Should be less than the
compliance level in each
frequency range.
b) Interference may occur in
the vicinity of equipment
marked with the following
symbol.