Duration of use
VivaSight 2 DLT can be used for up to 8 hours in total.
4.3. After Use
Visual Check
Examine the integrity of the product and check if there are any missing or broken parts. In case
any corrective actions are needed, act according to local hospital procedures.
Final Steps
1.
Disconnect the adapter cable from the Ambu displaying unit.
2.
Turn off the Ambu displaying unit by pressing the power button.
3.
VivaSight 2 DLT and the stylet are single-use devices and the adapter cable is for single-
patient use. Do not soak, rinse, sterilise or reuse the devices as this may leave
harmful residues or cause malfunction of the devices. The design and materials used
are not compatible with conventional cleaning and sterilisation procedures. 15
Disposal
VivaSight 2 DLT is a single use device and must be disposed of after use. VivaSight 2 adapter
cable is a single-patient use device and must be disposed of with its designated tube after use.
VivaSight 2 DLT and VivaSight 2 adapter cable, is considered contaminated after use and must
be disposed of in accordance with local guidelines for collection of infected medical devices
with electronic components.
5. Technical Product Specifications
5.1. Standards Applied
VivaSight 2 DLT function conforms with:
– EN 60601-1 Medical electrical equipment – Part 1: General requirements for basic safety and
essential performance.
– EN 60601-2-18 Medical electrical equipment – Part 2-18: Particular requirements for the basic
safety and essential performance of endoscopic equipment.
5.2. Specifications of VivaSight 2 DLT and Adapter Cable
VivaSight 2 DLT and
Adapter Cable
Power
Environmental Conditions
Ambient temperature
Ambient relative humidity
Ambient
atmospheric pressure
Storage recommendation
Specification
VivaSight 2 DLT and adapter cable can only be powered by
compatible Ambu displaying units. See section 2.2
Product Compatibility.
Operation
10 - 35 °C (50 - 95 °F)
30 - 75 %
70 - 106 kPa
-
14a
14b
Storage
10 - 25 °C (50 - 77 °F)
-
-
Store in a dry, cool and
dark place
11