WARNING: This is a controlled proprietary and confidential document.
AVISO: Este es un documento controlado, confidencial, y con derechos reservados.
7.
Slip the sterile sheath over the S
OUND
Catheter interconnect tab until the sheath is
appropriately seated, leaving the C
ARTO
connector uncovered.
8.
Inspect the SwiftLink* connector for damage.
Warning against possible patient injury!
Do not use the SwiftLink* connector if it
appears to be damaged. In this case, contact
your local Biosense Webster representative.
Warning against possible patient injury or
death!
Do not immerse the SwiftLink* connector in
fluid of any kind. If the SwiftLink* connector
is immersed in fluid, or damp, contact your
Biosense Webster representative.
Moisture trapped between the SwiftLink*
®
connector and the S
S
Catheter can
OUND
TAR
damage the SwiftLink* connector and/or the
®
S
S
Catheter, possibly causing
OUND
TAR
patient injury or death.
9.
Lift the lever on the SwiftLink* connector.
Slip the SwiftLink* connector on to the
®
S
S
Catheter interconnect tab until
OUND
TAR
the SwiftLink* connector is securely mated
®
with the S
S
Catheter handle.
OUND
TAR
Push the lever down, locking the
®
S
S
Catheter to the SwiftLink*
OUND
TAR
connector.
10. Carefully slip the sterile sheath over the
SwiftLink* connector. Cover enough of the
SwiftLink* connector cable such that it is
outside the sterile field.
11. Connect the other end of the SwiftLink*
connector to the ultrasound system. Ensure
that the ultrasound image appears in the
ultrasound system screen.
®
12. Connect the S
S
Catheter to the
OUND
TAR
Patient Interface Unit (PIU) via a sterile
extension cable, which has been connected
to the Q
port adaptor, as detailed in the
WIK
®
C
XP V9.7 System documentation
ARTO
(System Hardware section).
®
13. Connect the R
S
with Q
P
EF
TAR
WIK
External Reference Patch and ablation
catheter, if required, following the C
XP V9.7 System documentation (System
Hardware section).
DURING THE PROCEDURE
To conduct an ultrasound exam using the
®
S
S
Catheter:
OUND
TAR
1.
Create a vascular access with a catheter
introducer (hemostatic) large enough to
®
accommodate the S
S
Catheter
OUND
TAR
with heparinized saline.
2.
Before advancing or withdrawing the
®
S
S
Catheter, ensure that the
OUND
TAR
steering knobs are in the neutral position
and that the tension control knob is
released.
®
3.
Advance the S
S
Catheter into the
OUND
TAR
vasculature through the catheter introducer.
Fluoroscopy can aid in advancing the
catheter into the heart.
Warning against possible patient injury or
death!
Do not use excessive force to advance or
®
withdraw the S
S
Catheter.
OUND
TAR
Using excessive force can result in patient
injury or death.
To help prevent excessive force:
Ensure that both steering knobs are in
the neutral position, and that the tension
control knob is released before
advancing or withdrawing the
®
S
S
Catheter.
OUND
TAR
If you encounter strong resistance during
catheter navigation, discontinue the
procedure
Withdraw and redirect the S
OUND
Catheter as needed.
®
Manipulate the S
S
Catheter
OUND
TAR
carefully in order to avoid cardiac damage,
entanglement, perforation or tamponade
M-5276-646A
English
®
S
TAR
®
®
ATCH
®
ARTO
®
S
TAR
9
Release Date: 9/1/2009