4. Use of aScope 5 Broncho
Numbers in gray circles below refer to illustrations on page 2.
4.1. Preparation and inspection of aScope 5 Broncho
Lubricate the insertion cord with a water-based medical grade lubricant, to ensure the lowest
possible friction when the endoscope is inserted into the patient.
Visual inspection of the endoscope 1
1. Check that the pouch seal is intact. 1a
2. Make sure to remove the protective elements from the handle and from
the insertion cord. 1b
3. Check that there are no impurities or damage on the product such as rough surfaces,
sharp edges or protrusions which may harm the patient. 1c
4. Turn on the Ambu displaying unit. 2a 2b
Refer to the Ambu displaying unit Instructions for use for preparation and inspection
of the Ambu displaying unit. 2a 2b
Inspection of the image
1. Plug in the displaying unit connector into the corresponding connector on the
compatible displaying unit. Please ensure the colours are identical and be careful
to align the arrows. 3a 3b
2. Verify that a live video image appears on the screen by pointing the distal tip of the
endoscope towards an object, e.g. the palm of your hand. 4
3. Adjust the image preferences on the displaying unit if necessary (please refer to the Ambu
displaying unit Instructions for use).
4. If the object cannot be seen clearly, wipe the lens at the distal tip with a sterile cloth.
Preparation of aScope 5 Broncho
1. Carefully slide the control lever upwards and downwards to bend the bending section as
much as possible. Then slide the control lever slowly to its neutral position. Confirm that
the bending section functions smoothly and correctly. 5a
2. Carefully turn the rotation control ring left and right to rotate the insertion cord as much as
possible. Then turn the rotation control ring back to its neutral position. Confirm that the
rotation control ring functions smoothly and correctly. 5b
3. Press the endoscope buttons one after the other. Short press <1 second and long press
>1 second. For default setting please see displaying unit Instructions for use. 5c
10
Description
Electrical Safety Type
BF Applied Part.
Country of
manufacture:
Made in Malaysia.
UK Conformity
Assessed.
UK Responsible Person.
Symbols for the
aScope 5 Broncho
devices
Description
Medical Device.
UL Recognized
Component Mark
for Canada and the
United States.
Importer
(For products
imported into
Great Britain only).

Publicité

loading

Ce manuel est également adapté pour:

Ascope 5 broncho 2.2