Warning:
Portable and mobile radio communication devices can affect the correct function of this medical device.
Recommended separation distances for NON-LIFE SUPPORT patient equipment
CT1, CT2, CT3 microcellular telephones
DECT cellular telephones, wireless computer equipment
(modem, LAN)
Cellular, portable telephones (USA)
Cellular, portable telephones (e.g. GSM and NMT, EUROPA;
DECS 1800)
Walkie-talkie (rescue, police, fire dept., maintenance)
Cellular, case telephones
Mobiles (rescue, police, fire dept.)
For transmitters with frequencies lower than 800 MHz, the DISTANCE can be calculated using Equation A:
For transmitters with frequencies between 800 MHz and 2 GHz, the DISTANCE can be calculated using Equation B:
Where P is the nominal power of the transmitter in Watts (W) as declared by the transmitter manufacturer
All manuals and user guides at all-guides.com
RF source
d
Annex C
Typical Rated Current
(W)
0.01
0.25
0.6
2
8
5
16
100
d
4
P
2
3 .
P
31/01/2011 - 97050183-03-GB - 12
Distance
(m)
0.4
2
3
6
11
9
16
40