TROUBLESHOTING
FAULTS
Insufficient power or lack of
vibration
Insufficient liquid supply
None or insufficient lighting
All manuals and user guides at all-guides.com
PROBABLE CAUSES
The handpiece has not been properly
inserted on the coupling.
Power adjustment at the minimum
Selected operational mode not suitable
The insert has not been properly tightened
The insert is worn or deformed
Check and verify that the insert does not
show any micro-fractures in the threaded
area.
The device is damaged
The hose is damaged
The control microprocessor board is
damaged
Flow regulator closed
Tube obstructed
Hose is obstructed
Water circuit not operational
The lighting element is not assembled
The lighting element is counter polarized
Lighting element contacts are dusty
The lighting element is damaged
The hose is damaged
The control electronic board is damaged
INTERVENTION SUGGESTED
Verify connection
Verify the power settings on the
dental unit
Verify the operational mode on the
dental unit
Screw down using the key wrench
Substitute the insert
Substitute the insert
Substitute the handpiece
Request technical service assistance
Request technical service assistance
Verify the position of the flow
regulator on the instrument board or
the the hose fitting
Verify that the passage way hole in
the insert is not obstructed. If needed
use the syringe to blow air through it.
Request technical service assistance
Request technical service assistance
Verify the lighting element existence
Verify the light element tpolarity
Clean the lighting elements contacts
Substitute the lighting element
Request technical service assistance
Request technical service assistance