MAINTENANCE
At least once a week lubricate the O-Ring seals found in the handpiece coupling with silicone lubricating grease. Wearing
disposable gloves, apply a film of grease on the fingertips and lubricate with the fingers.
In the event that the spray is supplied with liquids that potentially leave incrustations, it is recommended that the circuit
be rinsed well with water at the end of each day's use.
The manufacturer requires the annual check or overhaul of the device on the part of authorised service assistance.
CLEANING, DISINFECTION AND STERILISATION
Before sanitisation operations, detach the handpiece from the micromotor and the motor from the hose.
To maintain a level of hygienic security, at the end of each use and within a short time, clean, disinfect and, if
applicable, sterilise the micromotor.
Electrical micromotor that may be sterilised are identified by the specific symbol found in the "Symbols" paragraph.
For cleaning and/or disinfection of the outside of the Electrical micromotor, use gauze or cotton soaked in ethyl alcohol
70% V.
Sterilization: steam autoclave at max. 135°C, 220 kPa (2.2 bar) for 5 minutes (250 cycles tested – EN ISO 11498).
Before
each
autoclave
"DAILY OIL", applying a quick squirt. Reassemble the motor on the hose and activate it so as to expel the excess
lubricant. Dry the outside of the Electrical micromotor with gauze or cotton wool before proceeding on with sterilisation.
 Do not use ultrasound cleaners.
 Do not immerse the handpiece in disinfectant or sanitizing solution.
 Never sterilise the handpiece in a dry heat steriliser.
 Do not leave the handpiece in the autoclave, but remove it immediately after the cycle.
 Periodically check the autoclave according to the requirements of the manufacturer.
Temperatures greater than the indicated limit may damage the micromotor.
TROUBLESHOOTING
FAULTS
The Electrical micromotor does
turn
The Electrical micromotor runs
but its speed is not adjustable
The Electrical micromotor
functions sporadically
The Electrical micromotor does
not provide sufficient torque
The Electrical micromotor
overheats
The Electrical micromotor turns
regularly The burr does not turn
The handpiece rotates on its
coupling.
The handpiece does not remain
inserted on the coupling.
Water leakage at the handpiece
Water leakage at the hose
coupling
Scrapping
Scrapping the device must be carried out with respect for the laws in force for electrical and electronic equipment,
according to the individual national laws.
The materials used for the manufacture are not harmful for human beings or animals that come into contact with or are
exposed to them.
All manuals and user guides at all-guides.com
cycle
lubricate
the
Electrical
PROBABLE CAUSES
Motor blocked
Blocked handpiece
Power supply broken
Hose broken
Adjustment command broken
Power supply malfunction
Instrument hose defective
Bearings damaged
Adjust the torque to its minimum
Endodontics mode selected
Power supply / motor malfunction
Insufficient cooling air.
Handpiece with strong friction
Handpiece not coupled properly
Handpiece broken
Defective transmission coupling
Blocked handpiece
Handpiece ruined
Motor hook ring ruined
O-ring seals worn or damaged
O-ring seals worn or damaged
micromotor
using
the
INTERVENTION SUGGESTED
Request technical service assistance
Lubricate the handpiece
Send the handpiece to technical
service assistance
Request technical service assistance
Request technical service assistance
Request technical service assistance
Request technical service assistance
Substitute the hose
Request technical service assistance
Recover the proper adjustment
Recover the proper adjustment
Request technical service assistance
Request technical service assistance
Lubricate the handpiece
Send the handpiece to technical
service assistance
Slide the handpiece out and reinsert
it properly.
Send the handpiece to technical
service assistance
Request technical service assistance
Send the handpiece to technical
service assistance
Substitute the handpiece
Request technical service assistance
Substitute the gasket
Substitute the gasket