CLEANING, DISINFECTING, STERILIZING
Warning!
The instrument is supplied not sterile.
Before use sterilize according to the above specifications.
Use gauze or cotton soaked in surgical alcohol to clean and disinfect syringe exterior.
Don't soak the syringe directly in solution. Don't use an ultrasonic cleaner.
In case of irregular water output, clean nozzle using the proper wire provided.
The external syringe casing made up of the nozzle and the grip (1 - Fig. 2) can be removed simply by pulling it off the
internal body (2 - Fig. 2) which remains connected to the cord) and sterilised in a steam autoclave at max.
135°C. Do not detach the nozzle from the grip for this operation.
In order to maintain a proper level of hygienic safety autoclave after the use on each patient.
Don't remove the nozzle from the handle for sterilisation purpose.
Remove the outside cover of the syringe from autoclave immediately after the cycle.
Warning!
Pay attention not to press A and B controls (Fig. 1) when removing and fitting again the
outside cover.
When fitting the outside cover rotate slowly into the position that allows a complete fitting until it clicks.
Warning!
Check the autoclave periodically according to the manufacturer's prescriptions!
Temperature exceeding the above stated limit may damage the handpiece!
Fig. 3 – Cleaning of the nozzle
Fig. 4 – Disinfecting of the syringe
1
Fig. 5 – Water filter
All manuals and user guides at all-guides.com
 - In the event of a poor delivery of air and/or water, do not try to
 - Whenever you note an appreciable decrease in the flow of water to
 - If a supplementary "Threesteril" syringe is present in the assistant's
 - In the event of water continuous leakage, turn the main switch of
2
TROUBLESHOOTING
adjust the pressure controls on the unit. It is possible to clean the
water duct on the nozzle (see fig. 3) by means of the supplied steel
wire. If a result is not achieved, ask for Technical Assistance.
the syringe, and in any case at least once every six months, you
must replace the filter (50) (2 – fig. 5) situated in the lower part
of the instrument tray.
tray, whenever you note an appreciable decrease in the flow of
water to the syringe you must replace the filter installed inside the
dentist unit body; this filter must in any case be replaced at least
once a year.
the unit off and ask for Technical Assistance.
31/01/2011 - 97050183-03-GB - 54