TECHNICAL FEATURES
Model
Transmission ratio
Side motor connection
UNI EN ISO 1797-1 Milling
cutters
Max milling cutter length (mm)
Maximum milling cutter
rotation speed
Optical Fibre
Warning
When using the R20-L adaptor for implant work, pay careful attention not to work when the tank is empty
in order to avoid the emission of air from the handpiece's external spray tube.
 Cool down operating field during use when necessary.
 If the air coolant flow is interrupted, immediately switch-off the handpiece.
 Do not connect or disconnect the handpiece from the micromotor until the micromotor has come to a
complete stop.
 Only operate the handpiece when the milling cutter is inserted.
 Do not use milling cutters with dimensions different than those indicated.
 Do not use worn or damaged milling cutters. Handle milling cutters with caution, wearing protective
gloves.
 Only use milling cutters (or other similar instruments with) that have a CE marking of conformity with
Directive 93/42/EEC.
 Before each use, check that the milling cutter is correctly locked.
 Do not exceed the speed recommended by the milling cutter's manufacturer.
 The milling cutter's lock button overheats if pushed during handpiece operation.
 Take care not to accidently press it against the patient's oral cavity.
 Do not use the instrument in the event of visible damage, the emission of unusual noises or vibrations,
abnormal overheating of if the milling cutter is not held firmly in place: Immediately stop use and
contact the technical assistance staff.
 When using this instrument it is recommended that a dental dam be used on the patient.
Fig. 1
Connection to Micromotor
Handpiece
Fig. 2 – Milling cutter insertion/
removal (straight handpiece)
All manuals and user guides at all-guides.com
"GOLDSPEED" HANDPIECES
S1
S1-L
1:1
1:1
ISO 3964
ISO 3964
Ø 2.35
Ø 2.35
latch bur
latch bur
26
26
50,000
50,000
-
Yes
HANDPIECE/MICROMOTOR CONNECTION/DISCONNECT
Handpieces without optical fibres
Insert handpiece and push it down until you hear a "click" Gently pull to detach it.
Handpieces with optical fibres
Engage handpiece, align the jutting handpiece base with the corresponding position on
the micromotor and push down until a "click" is heard. To detach handpiece, pull it
axially.
Do not connect or disconnect the handpiece from the micromotor until the
micromotor is comes to a complete stop.
MILLING CUTTER INSERTION/REMOVAL
Straight handpiece
To remove the milling cutter, rotate it in the direction of the arrow (Fig. 2 - A) until the
ring clicks (1). Insert the new milling cutter and bring the ring back to a closed
position with the two reference points aligned - (Fig. 2 - B).
Do not operate the motor if the ring is in the open position (reference points not
aligned).
M5-L
R20-L
1:5
20:1
ISO 3964
ISO 3964
Ø 1.6
Ø 2.35
latch bur
26
26
200,000
2,500
Yes
Yes
31/01/2011 - 97050183-03-GB - 62
E16
D1
16:1
1:1
ISO 3964
ISO 3964
Ø 2.35
Ø 2.35
latch bur
latch bur
22
26
3,125
50,000
600 (endo)
-
-