TECHNICAL FEATURES
Speed
Max torque
Driving air pressure (at the motor connection)
Water pressure
Max water flow
Spray air delivery
Intermittent operation
Note:
The rotation velocity measurement unit SI is "rad/sec".
Warning!
- Before every use, check that the handpiece is firmly secured to the air driven micromotor and that the burr is
correctly locked in the handpiece gripper.
- Castellini does not supply the burrs for the turbines and handpieces and PRESCRIBES the use of only burrs
(or other similar tools) that bear the EC mark of conformity with EEC Directive 93/42.
- Do not use the instrument if it is visibly damaged or if unusual noises and/or vibrations are produced.
Contact an authorised technical service centre.
- For the use of this instrument the rubber dam must be applied on the patient.
- Cool the operating area conveniently.
- Do not change the technical parametrer of the functionning
USING SAFELY: rules and recommendations
To ensure that the equipment is used safely, the user must abide by the set standards of hygiene and professional diligence.
The following points should also be kept in mind:
- During use, dust and fragments of material from the patient's mouth or the device being used may be thrust into the
surrounding environment (organic and inorganic particles, metal dust, liquids, potentially infected fluids and biological
materials):
PERSONNEL MUST DULY PROTECT THEIR EYES, BREATHING PASSAGES, MOUTH AND SKIN by wearing safety
glasses, face shields, masks and disposable gloves.
Operate the suction system at high speed in all operations likely to result in a discharge of materials, dust and aerosols to
minimise their dissemination.
N.B.: it is not advisable to use drinking water to supply the dental unit, since it may lead to the formation of a
biofilm in the waterlines and thus facilitate germ proliferation inside the unit itself; It is recommended to supply the unit
exclusively with dedicated liquids (Isotonic Saline Solution or I.P.Purified Water) using the Separate Supply system.
Fig. 1 – Anti-clockwise and power regulations
All manuals and user guides at all-guides.com
"AIR POWER 2" AIR DRIVEN MICROMOTOR
5.000 ÷ 20.000 ± 10% rpm
2,5 ± 10% N. cm
220÷260 kPa (2,2÷2,6 bar)
70 ÷ 140 kPa (0,7 ÷ 1,4 bar)
~ 70 cc/1 min at 0,9 bar
~ 6 lt/1 min
5 min work, 20 min stand-by
CONNECTION TO THE HOSE
The air driven micromotor has a four ways coupling to its hose (ISO 9168).
REGULATING DRIVING AIR PRESSURE
Connect the fitting with the pressure gauge (order code: L0001067) between
the tubing and the air driven micromotor. Operate the air driven micromotor
and check pressure.
AIR PRESSURE
Driving air pressure range: 220÷260 kPa (2.2÷2.6 bar).
REGULATING SPRAY WATER
Use the regulator on the under part of instruments table to control water
supply, (part. 8 – fig. 8 in chapter
To cut off the spray, adjust the controls on the dental unit.
1 rad/sec =
RPM
30
"Use and maintenance instructions
31/01/2011 - 97050183-03-GB - 64
").