Fig. 3
Milling cutter insertion/ removal
(handpiece counter-angle)
Fig. 4 – Lubrification
Warning:
Periodically check the autoclave, referencing the manufacturer's directions and recommendations.
Temperatures that exceed the recommended limits may damage the handpiece.
ANOMALIES
Handpiece not powerful, sudden decrease
in speed.
Noisy handpiece
Faulty insertion or removal of milling cutter Non-calibrated milling cutter diameter
Faulty spray release
Lack of light
Water leakage
All manuals and user guides at all-guides.com
Counter-angles
Lightly press your thumb at the centre of the button on the small head to remove the
milling cutter and insert a new one.
Milling cutter insertion
Milling cutter Ø 2.35: holding the button down, insert the milling cutter
completely and then slowly rotate in, pressing it down, until connection click is
heard; then release the button.
Milling cutter Ø 1.6: Insert the milling cutter completely holding down the
button and then release it. Make sure that the milling cutter stem is clean.
Pull on the milling cutter with your hand to make sure it has been properly inserted.
Do not use milling cutters that are longer than those recommended.
For proper locking, completely insert the milling cutter until it cannot go any further.
In order to avoid damage to the handpiece, do not activate the milling cutter release
mechanism until the milling cutter has come to a complete stop.
LUBRICATION
Lubricate the handpiece after each use and before and after each autoclave cycle using
the "DAILY OIL" aerosol spray.
Insert the applicator in the back part of the handpiece and release the spray until the
lubricant comes out of the head. Then, start-up the handpiece at slow speed with a
milling cutter inserted; gradually increase the speed and dry the excess lubricant.
Read all warnings and follow all instructions printed on the spray bottle label.
CLEANING, DISINFECTION AND STERILISATION
Warning: The instrument is not sterile when supplied.
Wear protective gloves. Remove milling cutter.
For external cleaning and/or disinfection of the handpiece use gauze or cotton saturated
with surgical alcohol. Do not use ultrasound cleaners.
The handpiece must not be immersed in disinfectant or sterilising liquids.
Clean the fibre optic terminals with alcohol-saturated gauze.
Do not use pointed or sharp instruments.
Clean the spray exit holes regularly with specially-provided probes.
In order to maintain the hygiene safety level, the handpiece must be sterilised in a
steam autoclave at max. 135°C, 220 kPa (2.2 bar) for 5 minutes (250 cycles tested – EN
ISO 7785-2) after use on each patient. Do not sterilise the handpiece in dry heat
sterilisers. It is important to lubricate the handpiece and cleanse the fibre optic terminals
after each autoclave cycle. Never leave the handpiece in the autoclave. It must be
removed immediately after each cycle.
PROBABLE CAUSES
Lack of lubrication
Worn rotating parts
Bent or very worn milling cutter
Lack of lubrication
Worn rotating parts
Worn milling cutter
Button not fully pressed down
Pincer worn
Obstructed nozzles
Bulb burnt in micromotor
Handpiece not properly inserted in
micromotor
O-Ring washers worn
RECOMMENDED INTERVENTION
Lubricate several times at intervals between
brief rotations.
Contact Technical Assistance
Insert a new milling cutter
Lubricate several times at intervals between
brief rotations.
Contact Technical Assistance
Use only high quality milling cutters (ISO
1797-1)
Use a new milling cutter
Full press down button in centre
Contact Technical Assistance
Check water regulator opening. Disconnect
handpiece, direct the cord towards basin,
check that water and air output.
Clean turbine using specially-provided steel
wire, taking care not to cause the ovalisation
of the nozzles.
Replace bulb
Properly insert handpiece
Replace and lubricate washers
31/01/2011 - 97050183-03-GB - 63