- Suction tubes
Cleaning and sanitizing:
- Cannulas connectors
Cleaning:
Disinfection
Sterilization:
- Suction system filter
Cleaning:
- Instruments
All manuals and user guides at all-guides.com
Aspirate the diluted solution of
then replace the disposable cannulas or rinse out and sterilize the reusable cannulas.
The Castellini company requires that only cannulas compliant with the 93/42/CEE directive
and subsequent amendments made – (ECC Directive 2007/47), be used. Replace the
disposable cannula.
Clean and decontaminate using Castellini
Umonium 38 Huckert's Int. or
Disinfect using
Rinse and sterilise the reusable cannulas in a steam autoclave at max. 135°C
Rinse and sterilise the connectors in a steam autoclave at max. 135°C
Clean and sanitize the conduits;
Activate the suction system by raising a cannula;
While the suction is on, remove the pressure plug;
Dispose of the residues and wash the filter under running water.
See the instructions for each instrument in Part
STER 3 PLUS
, according to the instructions provided;
STER 1 PLUS
.
ethyl alcohol 70% vol.
Operating Instruments
31/01/2011 - 97050183-03-GB - 31
– in this manual.