SAFETY SYSTEMS: SELF-DIAGNOSIS AND AUTOMATIC BLOCKING
The chair is equipped with a self-diagnosis system with warning signals.
When a button is pressed and the self-diagnosis system detects an anomaly, one of the following error
signals is displayed:
All the LEDs are flashing except for
1)
The backrest recline/return LEDs are flashing
2)
alternatively
Backrest recline/return LEDs are flashing simultaneously
3)
Chair raise/lowering LEDs are flashing simultaneously
4)
Chair raise LED flashes when the raise button is pressed
5)
Chair lowering LED flashes when the lower button is
6)
Backrest recline LED flashes when the recline button is
7)
Backrest return LED flashes when the return button is
8)
All manuals and user guides at all-guides.com
SIGNAL
P/S
pressed
pressed
pressed
REASON
Ophthalmic block active or instrument active
Assistant's table handle switch active
Backrest safety switch activated
Chair safety switch activated
Raise chair end of stroke
Lower chair end of stroke
Backrest recline end of stroke
Backrest return end of stroke