LCD displays the following message
TRANSMITTER HEAD
DISCONNECTED
SENSOR ERROR
PRESS A KEY
PLEASE WAIT
INCOMPLETE
ADJUSTMENT
REPEAT ADJUSTMENT
PRESS A KEY
PROGRAMMER FAILURE
SERVICE REQUIRED
NO LCD display; NO LED illumination
8
CODFR10065zz10svn.indd 8
CODFR10065zz10svn.indd 8
Cause
Transmitter is disconnected from the program unit
Technical problem with the transmitter's acoustic monitoring feature
The program unit requires 10 seconds between valve adjustment cycles
Adjustment cycle did not complete due to power interruption
Technical problem with the program unit
Power was not delivered to the program unit
Solution
Reconnect the transmitter to the program unit
Continued occurrence of this message might indicate a technical
problem; return for service.
Replace the transmitter.
OR
Use the packaged valve mode to adjust the valve. Press any key and
follow the instructions in Adjusting the Valve; Packaged Valve Mode
With an Implanted Valve.
If the message remains on the LCD longer than 10 seconds, return the
system for service.
Verify connection to power source. Verify power source is within
specification. Press any key and repeat the adjustment.
Persistent occurrence of this message might indicate a technical problem.
Return for service.
Return for service
1.
Ensure the correct type fuses are in place; replace if necessary
2.
Check power cord connection; check that power switch is on
3.
Check power supply integrity with a known functioning device
4.
Return for service