Télécharger Imprimer la page

Keysight Truevolt Serie Mode D'emploi page 93

Masquer les pouces Voir aussi pour Truevolt Serie:

Publicité

Les langues disponibles

Les langues disponibles

Features and Functions
Where:
Pre-Math - The measurement value before any math operations are done (including NULL).
l
Sensor - The raw sensor value; ohms for thermistor/RTD, volts for thermocouple (thermocouple meas-
l
urements apply to the 34465A and 34470A only).
Input/Ref - The DC signal voltage and the DC reference voltage measurements.
l
Peak - Displays a running history of the minimum peak, maximum peak, and peak-to-peak values of the
l
input signal. Peak measurements are high-speed (16 µs effective aperture) and are different from the
Min, Max, and Peak-to-Peak values gathered in Statistics.
1
After making one or more primary measurements for approximately 4 seconds, the DMM makes one sec-
ondary measurement.
For the Peak secondary measurement, the Clear Peaks softkey allows you to clear the accumulated his-
tory of the peak-to-peak function (resets the record of peak values).
94
Keysight Truevolt Series Operating and Service Guide

Publicité

loading

Ce manuel est également adapté pour:

Truevolt 34460aTruevolt 34461aTruevolt 34465aTruevolt 34470a