MEASure[:VOLTage]:{AC|DC}? [{<range>|AUTO|MIN|MAX|DEF} [, {<resolution>|MIN|MAX|DEF}]]
Sets all measurement parameters and trigger parameters to their
measurements and immediately triggers a measurement. The results are sent directly to the instrument's
output buffer.
The maximum range parameter (MAX) is 1000 V. However, the SAFETY LIMIT on the
front and rear HI/LO input terminals is 750 VAC (rms). The rms voltage is waveform
dependent. A sine wave is limited to 750 VAC (rms), but a 1000 Vpk square wave is
safe. Connection to AC mains power is further limited to CAT II (300 V). See
and Regulatory Information
operation of this instrument.
<range>: 100 mV, 1 V, 10 V, 100 V, 1000 V, AUTO (default) or DEFault
<resolution> (AC): optional and ignored; fixed at 6½ digits.
<resolution> (DC): See
NPLC.
The default is equivalent to 10 PLC.
Specify <resolution> in the measurement's units (V, A, Hz, Ω, and so on).
Configure DC voltage measurements using the 10 V range with 1 mV resolution.Then make and read one
measurement:
MEAS:VOLT:DC? 10,0.001
Typical Response:+8.54530000E+01
You can let autoranging select the measurement range, or you can manually select a fixed range.
l
Autoranging conveniently selects the range for each measurement based on the input signal. For fast-
est measurements, use manual ranging (autoranging may require additional time for range selection).
Autoranging (AUTO or DEFault), will generate an error if you specify a <resolution> because the instru-
l
ment cannot accurately resolve the integration time (especially if the input continuously changes). If
your application requires autoranging, specify DEFault for the <resolution> or omit the <resolution>
altogether.
Autoranging goes down a range at less than 10% of range and up a range at greater than 120% of
l
range.
If the input signal is greater than can be measured on the specified manual range, the instrument dis-
l
plays the word Overload on front panel and returns "9.9E37" from the remote interface.
Keysight Truevolt Series Operating and Service Guide
for more information on the safety features and the safe
Parameter
Resolution Table
Range, Resolution and
or
SCPI Programming Reference
default values
for AC or DC voltage
Safety
Typical Return
(none)
289