Gain Calibration Adjustment Overview
The ADC and zero calibration adjustment procedure must have been recently performed before beginning
gain calibration. Gain calibrations can be made using either the front or rear input (not available on the
34460A) terminals. On the 34461A/65A/70A, make sure the Front/Rear switch matches the terminals in
use.
Never turn off the instrument during a gain or flatness calibration. This may cause calibration memory for the
present function to be lost.
Valid Gain and Flatness Calibration Input Values
Gain calibration can be accomplished using the following input values.
Function
100 mV to 100 V
DC Volts
100 µA to 1 A
DC Current
Ohms (2W and 4W) 100 Ω to 100 MΩ
Frequency
100 µA to 1 A
[1]
AC Current (rms)
100 mV to 10 V
[1]
AC Volts (rms)
[1]
Valid frequencies are nominal frequencies ±1%.
Keysight Truevolt Series Operating and Service Guide
Range
Valid Amplitude Input Values
0.9 to 1.1 x Full Scale
1000 V
0.9 to 1.1 x Full Scale
3 A
10 A
0.9 to 1.1 x Full Scale
Auto
Input > 100 mV rms, 9.9 kHz to 110 kHz
0.9 to 1.1 x Full Scale
3 A
10 A
0.9 to 1.1 x Full Scale
100 V
750 V
Calibration Adjustment Procedures
450 to 1000 V
1.5 A to 2.5 A
1.8 A to 10 A
0.9 A to 2 A
2 A to 10 A
40 V to 110 V
400 to 750 V
529