Télécharger Imprimer la page

Keysight Truevolt Serie Mode D'emploi page 245

Masquer les pouces Voir aussi pour Truevolt Serie:

Publicité

Les langues disponibles

Les langues disponibles

SCPI Programming Reference
CONFigure:CONTinuity
Sets all measurement parameters and trigger parameters to their
urements.
Parameter
(none)
Configure the instrument for continuity measurements. Then make a measurement using an external trig-
ger with positive slope (rising edge) and read the measurement:
CONF:CONT
TRIG:SOUR EXT;SLOP POS
READ?
The range and resolution are fixed at 1 kΩ for continuity tests (a 2-wire resistance measurement).
l
The instrument beeps (if the beeper is enabled) for each measurement less than or equal to the con-
l
tinuity threshold (≤10 Ω), and the actual resistance measurement appears on the display.
From 10 Ω to 1.2 kΩ, the instrument displays the actual resistance measurement with no beep. Above
l
1.2 kΩ, the instrument displays "OPEN" with no beep.
The FETCh?, READ?, and
l
of its value.
Use
READ?
or
INITiate
l
246
MEASure:CONTinuity?
to start the measurement.
default values
Typical Return
+1.32130000E-02
queries return the measured resistance, regardless
Keysight Truevolt Series Operating and Service Guide
for continuity meas-

Publicité

loading

Ce manuel est également adapté pour:

Truevolt 34460aTruevolt 34461aTruevolt 34465aTruevolt 34470a