Features and Functions
Diode
This section describes how to configure diode tests from the front panel. The range and resolution are
fixed; the range is 10 VDC (with a 1 mA current source output).
Step 1: Configure the test leads as shown.
Step 2: Press
on the front panel to open a menu that specifies whether the DMM will beep to indic-
ate a successful diode test.
Diode measurements behave as follows:
0 to 5 V voltage displayed on the front panel, and instrument beeps when signal transitions into the 0.3
to 0.8 V threshold (if beeping is enabled)
> 5 V the front panel displays OPEN, and SCPI returns 9.9E37
78
Keysight Truevolt Series Operating and Service Guide