Measurement Tutorial
for high-frequency measurements, no anti-aliasing filter is provided in the multimeter. If you are con-
cerned about aliasing, you should add an external anti-aliasing filter.
Level Triggering
When digitizing, it is important to begin sampling at some defined point on the input signal such as when
the signal crosses zero volts or when it reaches the midpoint of its positive or negative peak amplitude.
Level triggering allows you to specify when (with respect to voltage and slope) to begin sampling. Refer
to
Level Triggering
for more information.
About Digitize Mode
Digitizing is available only for the 34465A/70A and requires the DIG option.
l
Digitizing allows you to take samples of the input signal, at a specified Sample Rate (for example, 50 kHz)
l
or Sample Interval (for example, 20 µS). You can specify the Duration as either an amount of Time or a
number of Readings (samples). You can use either auto, external, or level triggering. After pressing
[Acquire], press Acquire > Digitize. You can then select the digitizing/triggering parameters. After con-
figuring the digitizing parameters, press [Run/Stop]. Digitizing will begin when the specified trigger event
occurs.
Digitize mode is available only from the front panel and can only be used for DCV or DCI measurement
l
functions.
When you specify digitizing settings that conflict with present settings, the instrument displays a mes-
l
sage, and in most cases, adjusts settings to legal values. For example, if you configure digitizing for a
greater sample rate than can be achieved with the present integration time (NPLC setting), the instru-
ment displays a message and reduces the integration time. When you see a displayed message, such as
NPLC setting reduced to achieve digitize rate/interval, you can press Shift > Help > 1 View the last mes-
sage displayed > Select to view more information.
The DIG option increases the maximum reading rates from 5k readings/s (standard) to 50k readings/s
l
(maximum).
178
Keysight Truevolt Series Operating and Service Guide