LAN Services
LAN Services enables and disables the LAN services shown below.
After enabling/disabling one or more services, press Done > Apply Changes. After that, you must cycle
instrument power for the new settings to take effect.
The Web server enables or disables instrument programming from the instrument's Web interface.
The multicast DNS (mDNS) service is for use in networks where no conventional DNS server is installed.
Cycling power or resetting the LAN always enables mDNS.
The instrument Telnet port is 5024. Open SCPI sessions on the Telnet connection by entering:
telnet IP address 5024
Refer to the Keysight IO Libraries help for information on the VXI-11, Sockets, and HiSLIP protocols.
USB Settings
USB Settings configures the front panel USB (storage) and rear-panel USB (connectivity) connectors.
The SEC licensed option also gives you the ability to enable or disable the front panel USB interface via a
USB Front softkey. This option can be ordered as factory option or as customer-installable option. See
Models and Options
USB SCPI
USB SCPI enables or disables the rear-panel USB control port. After changing the interface state, cycle
instrument power to make the change take effect. When disabled, the interface cannot be configured by
the Keysight IO Libraries Connection Expert utility.
Keysight Truevolt Series Operating and Service Guide
for details. Without this option, the front panel USB interface is always enabled.