SCPI Programming Reference
OUTPut:TRIGger:SLOPe {POSitive|NEGative}
OUTPut:TRIGger:SLOPe?
Selects the slope of the voltmeter complete output signal on the rear-panel
Parameter
{POSitive|NEGative}. Default: NEG
Configure DC voltage measurements and make two measurements. The signal on the rear-panel VM
Comp connector outputs a positive pulse as each measurement is completed:
CONF:VOLT:DC 10
SAMP:COUN 2
OUTP:TRIG:SLOP POS
INIT
On the 34460A, this requires option 34460A-LAN or option 3446LANU.
l
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
204
VM Comp
Typical Return
POS or NEG
Keysight Truevolt Series Operating and Service Guide
BNC connector.