[SENSe:]CURRent:AC:SECondary {"OFF"|"CALCulate:DATA"|"FREQuency"|"CURRent[:DC]"}
[SENSe:]CURRent:AC:SECondary?
Selects the secondary measurement function for AC current measurements.
{"OFF"|"CALCulate:DATA"|"FREQuency"|"CURRent[:DC]"},
default "OFF"
Configure AC current measurements using the 1 A range. Select frequency as the secondary
measurement. Make and retrieve measurements.
CONF:CURR:AC 1
CURR:AC:SEC "FREQ"
READ?;DATA2?
Typical Response: +0.50000000E+00;+1.00000000E+03
"CALCulate:DATA" - (Applies only to the 34465A and 34470A.) The measurement value before any
l
math operations are done (including NULL).
"FREQuency" - The frequency measurement of the input signal.
l
"CURRent[:DC]" – The DC current measurement of the input signal. Applies only when using the
l
instrument from the front panel. Setting is ignored for measurements made from a remote interface.
If you require both AC and DC current measurements from remote, program each function separately,
see CONFigure:CURRent:{AC|DC}.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
[SENSe:]CURRent[:DC]:APERture {<seconds>|MIN|MAX|DEF}
[SENSe:]CURRent[:DC]:APERture? [{MIN|MAX|DEF}]
Specifies the integration time in seconds (called aperture time) for dc current measurements.
Use this command for precise control of the DMM's integration time. Use NPLC
(see
acteristics (NPLC > 1).
(Without the DIG option) 200 µs to 1 s (2 µs resolution), default: 100 ms. (With the
DIG option) 20 µs to 1 s (2 µs resolution), default: 100 ms.
Enable aperture time mode, set the aperture time to 300 ms:
CURR:APER:ENAB ON
CURR:DC:APER 300E-03
Keysight Truevolt Series Operating and Service Guide
Parameter
[SENSe:]CURRent[:DC]:NPLC
Parameter
SCPI Programming Reference
Typical Return
"CALC:DATA"
) for better power-line noise rejection char-
Typical Return
+1.00000000E-
01