SCPI Programming Reference
[SENSe:]{FRESistance|RESistance}:SECondary {"OFF"|"CALCulate:DATA"}
[SENSe:]{FRESistance|RESistance}:SECondary?
Selects the secondary measurement function for all resistance measurements.
Parameter
{"OFF"|"CALCulate:DATA"}, default "OFF"
Configure resistance and secondary measurements. Make resistance measurement and retrieve sec-
ondary measurement.
CONF:RES AUTO,MAX
RES:SEC "CALC:DATA"
READ?;DATA2?
Typical Response: +3.01566373E-10;,+9.91000000E+3
"CALCulate:DATA" - (Applies only to the 34465A and 34470A.) The measurement value before any
l
math operations are done (including NULL).
Use
READ?
or
INITiate
l
measurement.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
356
to start the measurement. Use
Typical Return
"CALC:DATA"
[SENSe:]DATA2?
to retrieve the secondary
Keysight Truevolt Series Operating and Service Guide