3 A and 10 A Current Path Fuse Replacement
This section describes the procedure for testing and replacing the 3 A and 10 A current path fuses inside
the instrument.
Only qualified, service-trained personnel who are aware of the hazards involved
should remove instrument covers. Always disconnect the power cable and any
external circuits before removing the instrument cover. Some circuits are active and
have power applied even when the power switch is turned off.
Always disconnect all inputs, cords, and cables before disassembling the instru-
ment.
Do not turn the screw that fastens the shield in place, as this will affect the instru-
ment's calibration.
Tools Required
T20 Torx driver (most disassembly)
l
Small, flat-blade screw driver (for prying)
l
Keysight Truevolt Series Operating and Service Guide
Service and Repair
495