SCPI Programming Reference
CALCulate:TRANsform:HISTogram:RANGe:AUTO {ON|1|OFF|0}
CALCulate:TRANsform:HISTogram:RANGe:AUTO?
Enables or disables automatic selection of the histogram's lower and upper range values.
Parameter
{ON|1|OFF|0}. Default: ON. 0 (OFF) or 1 (ON)
See
.
Example
ON: the instrument uses the first 1,000 measurements to set the lower and upper range values.
l
OFF: the lower and upper range values are specified by
l
{LOWer|UPPer}.
Setting the lower or upper range value (CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer) dis-
l
ables automatic selection of the lower and upper range values
late:TRANsform:HISTogram:RANGe:AUTO
The instrument restarts automatic range value selection (if enabled) when INITiate,
l
READ?
is executed.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer} {<value>|MIN|MAX|DEF}
CALCulate:TRANsform:HISTogram:RANGe:{LOWer|UPPer}? [{MIN|MAX|DEF}]
Sets the histogram's lower and upper range values. Setting the lower or upper range value
late:TRANsform:HISTogram:RANGe:{LOWer|UPPer) disables automatic selection of the lower and upper
range values
(CALCulate:TRANsform:HISTogram:RANGe:AUTO
Parameter
-1.0E+15 to -1.0E-15,
+1.00000000E+06
or
0.0 (default)
or
+1.0E-15 to 1.0E+15
See
.
Example
If automatic range value selection is enabled
l
the query returns the computed range value. If no histogram data exists, 9.91E37 (Not a Number) is
returned.
Lower and upper range values are computed automatically if
l
late:TRANsform:HISTogram:RANGe:AUTO
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
220
Typical Return
OFF).
Typical Return
(CALCulate:TRANsform:HISTogram:RANGe:AUTO
is ON.
CALCulate:TRANsform:HISTogram:RANGe:
(CALCu-
OFF).
CALCu-
Keysight Truevolt Series Operating and Service Guide