DC Current 10 A Gain Calibration
This calibration procedure applies to the 34461A, 34465A, and 34470A only (not applicable to the
34460A).
Review
Test Considerations
Configuration: DC Current
1. Configure the 10 A range and set the integration time to 100 PLC.
2. Apply the 5 A input.
3. Enter the actual applied input current as the Cal Value (see
Step Succeeded message indicates success; if the display shows Calibration Step Failed, check the
input value, range, function, and entered calibration value and repeat the calibration step.
4. Store the new calibration constants (see
This calibration takes less than 4 seconds to complete.
Keysight Truevolt Series Operating and Service Guide
and
Flatness Calibration Considerations
Storing the Calibration
Calibration Adjustment Procedures
before performing any calibration.
Entering Calibration
Values).A Calibration
Constants).
Next Procedure
569