SCPI Programming Reference
SYSTem:ACALibration:TIME?
Returns the time of the last autocalibration in the format hh,mm,ss.sss.
Parameter Typical Return
(none)
20,15,30.000
Return the calibration time:
SYST:ACAL:TIME?
(34465A/70A only)
l
The date is based on the instrument's real-time clock. Set the date for the instrument's real-time
l
clock with SYSTem:DATE.
The real-time clock does not adjust itself for time zone changes or daylight savings time. Be sure to
l
set the date and time when you first receive your instrument.
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
Use
*CAL?
to perform autocal.
l
See Also
SYSTem:TIME
SYSTem:BEEPer[:IMMediate]
Issues a single beep.
Parameter Typical Return
(none)
(none)
Issue a single beep:
SYST:BEEP
Programmed beeps may be useful for program development and troubleshooting.
l
You can issue a beep with this command regardless of the current beeper state
l
:STATe).