SCPI Programming Reference
MMEMory:FORMat:READing:INFormation {OFF|ON}
MMEMory:FORMat:READing:INFormation?
Disables/enables header information and reading numbers in stored data.
Parameter
{OFF|ON}. Default OFF.
Enable the format for the .csv file:
MMEM:FORMat:READing:INFormation ON
Applies only to the 34465A and 34470A.
l
The format for the .csv file
l
34465A/34470A, are affected as follows:
OFF = one reading per row, no other header or reading information.
ON = enables a file header and reading numbers. The file header contains the start date and start
time of the first reading and, if the sample source is timer (SAMPle:SOURce:TIMer), the sample inter-
val. Each data row includes a reading number and the reading.
Affects the .dat (binary) file created by
l
OFF = one reading per row, no other header or per reading information.
ON = one time stamp row consisting of eight 16-bit integers for: year, month, day of week, day, hour,
minute, and milliseconds. 32-bit integer reading number and 64-bit IEEE-754 reading.
MMEMory:FORMat:READing:RLIMit {OFF|ON}
MMEMory:FORMat:READing:RLIMit?
When enabled, limits the maximum number rows per stored .csv data file to one million, including any
header row, if enabled.
Parameter
{OFF|ON}. Default ON.
Disable row limits for the .csv file:
MMEM:FORM:READ:RLIMit OFF
Applies only to the 34465A and 34470A.
l
Affects the format for the .csv file created by the
l
logging/digitizing features in the 34465A/34470A.
This feature enables you to accommodate common spreadsheet, database and data analysis pro-
l
grams that have limitations of 1 million rows per file.
ON - Readings are stored in files named dat00001.csv, dat00002.csv, dat00003.csv ...
l
OFF - Readings are is stored in a single file named dat00001.csv. The file size is limited to 2
l
4.294967296 GBytes (the maximum allowed by the FAT32 file system).
304
OFF or ON
(MMEMory:STORe:DATA
MMEMory:STORe:DATA
OFF or ON
MMEMory:STORe:DATA
Typical Return
), or by the data logging/digitizing features in the
as follows:
Typical Return
command, or by the data
Keysight Truevolt Series Operating and Service Guide
32
=

Publicité

loading

Ce manuel est également adapté pour:

Truevolt 34460aTruevolt 34461aTruevolt 34465aTruevolt 34470a