SCPI Programming Reference
CONFigure[:VOLTage][:DC]:RATio [{<range>|AUTO|MIN|MAX|DEF} [, {<resolution>|MIN|MAX|DEF}]]
Sets all measurement and trigger parameters to their
measurements. Also specifies the range and resolution.
To calculate a ratio, the instrument measures a DC reference voltage applied to the Sense terminal and
the DC signal voltage applied to the Input terminals. It then uses the formula:
     Ratio = DC signal voltage / DC reference voltage
<range>: {100 mV|1 V|10 V|100 V|1000 V}. Default: AUTO (autorange).
< resolution >: See
Resolution Table
to 10 PLC.
Configure the instrument for DC voltage ratio measurements using the 100 V range with 1 mV resolution.
Then make a measurement using an external trigger with positive slope (rising edge) and read the meas-
urement:
CONF:VOLT:DC:RAT 100,0.001
TRIG:SOUR EXT;SLOP POS
INIT
FETC?
Typical Response: +4.27150000E+00
The Sense terminals have a maximum measurable input of ±12 VDC. Autoranging is automatically
l
selected for reference voltage measurements on the Sense terminals.
The Input LO and Sense LO terminals must have a common reference and cannot have a voltage dif-
l
ference greater than ±2 V.
The specified measurement range applies only to the signal connected to the Input terminals. The sig-
l
nal on the Input terminals can be any DC voltage up to 1,000 V.
For the Input terminals, you can allow the instrument to select the measurement range by autor-
l
anging or you can select a fixed range using manual ranging. Autoranging decides which range to use
for each measurement based on the input signal. For faster measurements, use manual ranging on
each measurement (additional time is required for autoranging to select a range).
Autoranging (AUTO or DEFault), will generate an error if you specify a <resolution> because the instru-
l
ment cannot accurately resolve the integration time (especially if the input continuously changes). If
your application requires autoranging, specify DEFault for the <resolution> or omit the <resolution>
altogether.
Autoranging goes down a range at less than 10% of range and up a range at greater than 120% of
l
range.
If the input signal is greater than can be measured on the specified manual range, the instrument dis-
l
plays the word Overload on front panel and returns "9.9E37" from the remote interface.
Use
READ?
or
INITiate
l
See Also
CONFigure?
254
Parameter
or
Range, Resolution and NPLC.
to start the measurement.
default values
for DC voltage to DC voltage ratio
The default is equivalent
Keysight Truevolt Series Operating and Service Guide
Typical
Return
(none)